1. Basic Structure and Function of an Implantable Cardioverter Defibrillator (ICD)
An ICD is a small device implanted in the chest that monitors the heart's rhythm and delivers electrical shocks when it detects dangerous arrhythmias, such as ventricular tachycardia (VT) or ventricular fibrillation (VF). The device consists of two main components: a pulse generator and one or more electrode leads. The pulse generator houses the battery, sensing and pacing circuits, and a capacitor. The battery provides the energy needed for the device to function, while the capacitor is responsible for storing and discharging electrical energy during a shock. The sensing and pacing circuits continuously monitor the heart’s electrical activity, detect abnormal rhythms, and deliver pacing pulses if necessary.
Early ICDs used epicardial electrodes, which required open-heart surgery for placement. Over time, this evolved into subcutaneous electrodes, and now most devices use transvenous endocardial leads, which are inserted through a vein and placed directly in the heart. This has greatly simplified the implantation process. The electrode leads transmit signals from the heart to the pulse generator and also deliver pacing or defibrillation shocks to the heart. Depending on the electrode configuration, the shock can be delivered either through the transvenous lead or via the pulse generator housing itself.
The primary function of an ICD is to recognize and treat both tachyarrhythmias and bradycardia. Its ability to manage bradycardia is similar to that of a pacemaker. However, the focus here will be on how the ICD identifies and treats rapid heart rhythms.
1.1 Recognition of Rapid Arrhythmias
The ICD uses several criteria to identify tachyarrhythmias, including heart rate (or its corresponding R-R interval), duration, and additional features like sudden onset and stability. These criteria are applied in two stages: initial recognition and re-identification. Initial recognition is used to detect the first occurrence of an arrhythmia, while re-identification helps confirm that the arrhythmia persists after initial treatment attempts.
VT and VF are primarily distinguished by their heart rates. Different manufacturers may use different terminology to express these criteria. For example, some companies like CPI may specify frequency in beats per minute (bpm) and duration in seconds, while others like Medtronic might use R-R intervals in milliseconds and the number of cardiac cycles. Understanding these differences is crucial for proper programming of the device.
For instance, if the VT recognition threshold is set at 150 bpm with a duration of 10 seconds, the ICD would initiate treatment once these conditions are met. In the case of a CPI Ventak PRx III device, the duration starts counting only after the heart rate meets the set threshold. The ICD continuously monitors each heartbeat, comparing the R-R interval to the pre-set criteria. If eight out of ten consecutive heartbeats meet the criteria, the ICD confirms the diagnosis and begins timing the duration.
Similarly, for Medtronic devices, the same scenario might require setting the R-R interval to 400 ms and the duration to 24 cardiac cycles (which equates to approximately 10 seconds). The difference lies in how the ICD calculates the start of the duration—CPI starts timing after meeting the rate, while Medtronic may consider both rate and cycle count simultaneously.
In addition to heart rate and duration, the ICD may also use other features like sudden onset and stability to differentiate between VT and other types of tachycardia, such as sinus tachycardia or atrial fibrillation. Sudden onset refers to a rapid change in heart rate, typically defined as a percentage decrease in the R-R interval compared to normal sinus rhythm. Stability, on the other hand, measures the regularity of the rhythm, with VT usually being more regular than atrial fibrillation.
1.2 Treatment of Rapid Arrhythmias
Once a tachyarrhythmia is detected, the ICD can respond using two main methods: electric shock (defibrillation) or anti-tachycardia pacing (ATP).
(1) Electric Shock: Most modern ICDs can deliver shocks ranging from 0.1 to 34 joules. Some models can go up to 42 joules, and many can deliver multiple shocks in succession. Both VT and VF can be treated with this method. The first shock is typically set slightly higher than the patient’s measured defibrillation threshold, and subsequent shocks may use maximum energy to ensure effectiveness.
(2) ATP: This technique involves delivering a series of rapid pacing impulses to interrupt the arrhythmia. There are two common forms: short-array rapid pacing and ramp pacing. Short-array pacing uses a pacing rate that is slightly faster than the tachycardia, often set at 75–80% of the tachycardia cycle length. Ramp pacing gradually decreases the pacing rate over time, which can be more effective in certain cases but carries a higher risk of accelerating the arrhythmia if not properly managed.
While VF must be treated with electric shock, VT can be addressed with either ATP or shock therapy. ATP is often preferred as it is less invasive and can be more effective in terminating the arrhythmia without causing discomfort. If ATP fails, low-energy shocks may be used first, followed by high-energy shocks if necessary.
In addition to treating arrhythmias, ICDs also record detailed information about each event, including the ECG waveform, heart rate, R-R intervals, treatment mode, and response. This data is invaluable for follow-up and adjusting the device settings to improve future outcomes. Many ICDs also have built-in electrophysiology testing capabilities, allowing doctors to induce arrhythmias and test the device’s response during follow-up visits.